Clamp Meter F01 Clamp meter with real-time RMS measuring / AC measuring / voltage measuring display-hold-function / automatic shutdown function / high voltage warning
|
The Clamp Meter F01 is designed for easy measuring of alternating current. In addition, the clamp meter can also determine the AC/DC voltage. Further functions of the clamp meter are resistance measurements and continuity testing. When high voltage is detected, the clamp meter will give out a warning. The clamp has a diameter of 26mm. The power supply of the Clamp Meter F01 is provided by a 9V battery. With this, the clamp meter can perform up to 25,000 measurements. Finally, the automatic shutdown function is another high quality feature of the clamp meter. If there are any questions about this Clamp Meter F01 have a look at the technical specification below or call: UK customers +44(0) 23 809 870 30 / US customers +1-410-387-7703. Our technicians and engineers will gladly advise you in all clamp meters and all products in the field of measuring instruments, regulation and control and scales.
|
|
- RMS measuring - High voltage warning - AC measuring (A, V)
|
- Continuity tester - Resistance measuring - Automatic shutdown function
|
Technical Specifications of the Clamp Meter F01
|
AC A measurements
|
Range
|
Resolution
|
Accuracy
|
0.2 A … 39.99 A
|
10 mA
|
± 1.5 % + 10 digits
|
40 A … 399.9 A
|
0.1 A
|
± 1.5 % + 2 digits
|
400 A … 600 A
|
1 A
|
± 1.5 % + 2 digits
|
AC / DC V measurements
|
|
|
Range
|
Resolution
|
Accuracy
|
0.2 V … 39.99 V
|
10 mV
|
± 1 % + 5 digits
|
40 V … 399.9 V
|
0.1 V
|
± 1 % + 2 digits
|
400 V … 600 V
|
1 V
|
± 1 % + 2 digits
|
Electrical resistance measurements
|
|
|
Range
|
Resolution
|
Accuracy
|
0 Ω … 399.9 Ω
|
0.1 Ω
|
± 1 % + 3 digits
|
Continuity testing
|
|
|
Acoustic signal
|
R < 40 Ω
|
|
General Technical data of the Clamp Meter F01
|
Display
|
LCD with 4,000 digits
|
Clamp diameter
|
< 26 mm
|
Power supply
|
9 V battery
|
Operating duration
|
25,000 measurements at 10 seconds
|
Dimensions
|
70 x 193 x 37 mm
|
Weight
|
260 g
|
Protection class
|
IP40
|
Operating conditions
|
0 … 50 °C 0 … 90 % humidity
|
Standards
|
EN 61010-1 EN 61010-2-032 Installation category: III Operating voltage: 600 V
|
Electromagnetic compatibility
|
EN 61326-1
|
Delivery contents of the Clamp Meter F01
|
1 x Clamp Meter F01, 1 x set of measuring lines, 1 x carrying case, 1 x 9 V battery, 1 x user manual
|
Here you can find further similar products regarding 'clamp meters':
|
- Clamp Meter F03 (clamp meter AC / DC voltage and resistance measurement)
- Clamp Meter F05 (clamp meter for single-phase performance measurement, measurement of starting current)
- Clamp Meter F07 (clamp meter with inrush measurement, performance measurement and measuring adapter input)
- Clamp Meter F09 (clamp meter for measurement of direct current and alternating current, performance measurement)
More clamp meters of PCE Instruments:
- PCE-DC 1 Clamp Meter (Clamp meters up to 200 A AC/DC, frequency measurement included, voltage surge category III)
- PCE-DC 2 Clamp Meter (Clamp meters up to 200 A AC/DC, ohms measurement included, voltage surge category III)
- PCE-DC 3 Clamp Meter (Clamp meters up to 80 Amp AC/DC with high resolution, category III surge protection)
- PCE-LCT 1 Clamp Meter (Clamp meters up to 100 A AC, category III)
|
|
|
Here you will find an overview of all the measuring instruments available at PCE Instruments. |
Contact: PCE Instruments UK Limited
Unit 11 Southpoint Business Park
Ensign Way, Southampton
United Kingdom, SO31 4RF
Phone: +44(0) 23 809 870 30
Fax: +44(0) 23 809 870 39
|
Contact: PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8
Jupiter 33458 FL
USA
Phone: +1-410-387-7703
Fax: +1-410-387-7714
|