PCE-LCT 1 leakage current meter digital leakage current meter to test leakage current / RMS measurement / continuity and resistance test / measures voltage up to 400 V
|
With PCE-LCT 1 leakage current meter it is possible to measure leakage current with a range up to 100 A. The PCE-LCT 1 leakage current meter can measure leakage current while current flows through electrical conductors that are not designed for conduction current. The PCE-LCT 1 leakage current meter can detect leakage currents and show them on the display. With the help of the PCE-LCT 1 leakage current meter, you can detect insulation failures in places where machines cannot be disconnected but the insulation still needs to be checked. The PCE-LCT 1 leakage current meter measures current flow, voltage up to 400 V, resistance and continuity. The PCE-LCT 1 is an ideal instrument for electricians and meets VDE standards. The PCE-LCT 1 has a compact design and can be placed in a pocket. The rubber protection protects the instrument against being damaged, so it is ideal for being used in electrical inspections and in workshops. Here you will find an overview of all the leakage current meters available at PCE Instruments.
|
View or print user manual of the PCE-LCT 1 leakage current meter
|
- Measures leakage current - Measures current up to 100 A - Measures voltage up to 400 V - Continuity test - Measures resistance up to 400 Ω
|
- Digital bar graph - High resolution - Max./min. values are viewed on the display - Rubber protector - Test cables are included
|
Technical specifications |
Leakage current and current (AC A) |
Measurement ranges |
Resolution |
Accuracy |
50/60 Hz |
400 Hz ... 1 kHz |
40 mA |
10 µA |
± (1 % + 0.5 mA) |
± (4.5 % + 0.5 mA) |
400 mA |
100 µA |
± (3 % + 5 mA) |
± (3 % + 5 mA) |
4 A |
1 mA |
± (4 % + 0.1 mA) |
± (4 % + 0.1 A) |
40 A |
10 mA |
± (4 % + 1 A) |
± (4 % + 1 A) |
80 A |
100 mA |
± (2.5 % + 1 A) |
± (3 % + 1 A) |
80 A - 100 A |
100 mA |
± (5 % + 1 A) |
± (5 % + 1 A) |
Voltage (AC V) |
Measurement ranges |
Resolution |
Accuracy |
50/60 Hz |
400 Hz ... 1 kHz |
400 V |
0.1 V |
± (1.5 % + 2 digits) |
± (2 % + 4 digits) |
Resistance |
Measurement ranges |
Resolution |
Accuracy |
Alarm |
400 Ω |
0.1 Ω |
± (1 % + 2 digits) |
Measurement value <38Ω |
General technical specifications |
Display |
3 3/4 |
Measured electrical properties |
AC V, AC A, resistance |
Security standards |
CAT III 300 V/CAT II 600 V |
Diameter of the clip |
30 mm |
Input impedance |
10 MΩ |
Registration rate |
2 Hz (numerical values), 20 Hz (bar graph) |
Operating temperature |
-10 ºC ... 50 ºC |
Operating humidity |
< 85 % H.r. |
Storing temperature |
-20 ºC ... 60 ºC |
Auto shut-off |
after 30 minutes |
Power |
9 V battery |
Input current |
10 mA |
Weight |
200 g |
Dimensions |
210 x 62 x 35.6 mm |
Delivery contents 1 x PCE-LCT 1 leakage current meter, 1 x 9 V battery, 1 x test cable, 1 x carrying case, user manual
|
|
|
|
Here you will find an overview of all the measuring instruments available at PCE Instruments.
|
Contact: PCE Instruments UK Limited
Unit 11 Southpoint Business Park
Ensign Way, Southampton
United Kingdom, SO31 4RF
Phone: +44(0) 23 809 870 30
Fax: +44(0) 23 809 870 39
|
Contact: PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8
Jupiter 33458 FL
USA
Phone: +1-410-387-7703
Fax: +1-410-387-7714
|
Below you will find an overview of the different product groups of measuring instruments
|
|
|
|
|
|