PCE-VS Rotameter plastic rotameter for measuring flow in pipes and vents for air Nm3/h or water l / h , depending on model
|
The rotameter is an instrument to determine the flow of fluids, e.g. liquids or gases in pipes. The rotameter of the VS series operates on the principle of the body in suspension. In the rotameter's conical plastic test tube there is a body in suspension that is lifted by the inertia of the fluid itself. How high the body in suspension reaches depends on the flow rotameter. If you increase the flow, the flow resistance increases. The suspension of the rotameter body rises and increases the body surface between the suspension and the glass tube. This decreases the flow resistance until it is identical with the sum of the strength of the suspended body weight and buoyancy. The level serves as a measure for the flow and can be read on a scale on the rotameter. Therefore, the rotameter is available in different versions. The rotameter is sent by default with a scale in l / h for water or Nm ³ / h for air. The reading of the flow is facilitated by two adjustable set-point indicators integrated into the rotameter. For more information on rotameters that detect flow please click here.
|
|
Rotameter series PCE-VS
|
- Ability to read l / h Nm3 / h
- High strength
- Corrosion resistant
- High repeatability
- High reproducibility
- Range of 1.5 - 60,000 l / h
|
- Easy installation
- No need for any external power
- Accuracy class 4, according to VDI 3513
- For liquids and gases
- Setting the theoretical value indicators
- Floating body of PVDF (polyvinylidene fluoride)
|
Technical specifications
|
Series |
VS11... |
VS12... |
VS13... |
|
|
VS32... |
VS33... |
Specificactions |
Accuracy |
- Accuracy class 4, according to VDI 3513 |
Nominal precision 30 ºC |
PN 10 |
Medium temp. |
0 ... 60 ºC |
0 ... 100 ºC |
0 ... 40 ºC |
Measurement tube |
PA (Trogramid) |
PSU (Polisuifon) |
PVC |
Floating body |
PVDF (Polyvinylidene fluoride) |
Ring O |
EPDM (Ethylene Propylene Diene Type M ASTM ) |
Series VS1... (length = 335 mm) |
Type |
Nominal value |
Range for water |
Range for air 0 ... 10 bar |
VS 1_25A... |
DN25 |
50 - 500 l/h |
1.5 - 42 Nm3/h |
VS 1_25B... |
DN25 |
100 - 1,000 l/h |
2.5 - 87 Nm3/h |
VS 1_32C... |
DN32 |
150 - 1,500 l/h |
4 - 135 Nm3/h |
VS 1_32E... |
DN32 |
250 - 2,500 l/h |
7 - 237 Nm3/h |
VS 1_40D... |
DN40 |
200 - 2,000 l/h |
6 - 174 Nm3/h |
VS 1_40F... |
DN40 |
300 - 3,000 l/h |
9 - 324 Nm3/h |
VS 1_40G... |
DN40 |
600 - 6,000 l/h |
17 - 522 Nm3/h |
VS 1_50G... |
DN50 |
600 - 6,000 l/h |
17 - 525 Nm3/h |
VS 1_50H... |
DN50 |
1,000 - 10,000 l/h |
29 - 903 Nm3/h |
VS 1_50I... |
DN50 |
1,500 - 15,000 l/h |
53 - 1215 Nm3/h |
VS 1_65J... |
DN65 |
2,000 - 20,000 l/h |
55 - 1635 Nm3/h |
VS 1_65K... |
DN65 |
3,000 - 30,000 l/h |
80 - 2274 Nm3/h |
VS 1_65L... |
DN65 |
8,000 - 80,000 l/h |
- - - |
Warnings: Note the different ranges of precision air. The precise ranges can be found here
|
Dimensions VS1... (L = 335 mm)
|
|
Installation and assembly indications for the PCE-VS Rotameter Rotameters should be installed vertically and free of tension. It is necessary to provide a route of entry and exit, input of approx. 10 x DN, output of approx. 5 x DN
|
|
G
|
DN
|
L
|
L1
|
L2
|
ØD
|
1 1/2"
|
25
|
335
|
341
|
385
|
60
|
2"
|
32
|
335
|
341
|
393
|
72
|
2 1/4"
|
40
|
335
|
341
|
403
|
83
|
2 3/4"
|
50
|
335
|
339
|
417
|
103
|
3 1/2"
|
65
|
335
|
341
|
429
|
122
|
|
Series VS3... (length = 165...200 mm)
|
Type
|
Nominal value
|
Water range
|
Air range 0 ... 10 bar
|
VS3_10P...
|
DN10
|
1.5 - 15 l/h
|
0.01 - 1.8 Nm3/h
|
VS3_10Q...
|
DN10
|
2.5 - 25 l/h
|
0.2 - 3.2 Nm3/h
|
VS3_10R...
|
DN10
|
5 - 50 l/h
|
0.5 - 6.4 Nm3/h
|
VS3_10T...
|
DN10
|
10 - 100 l/h
|
0.8 - 10 Nm3/h
|
VS3_15S...
|
DN15
|
8 - 80 l/h
|
0.6 - 9.5 Nm3/h
|
VS3_15U...
|
DN15
|
15 - 150 l/h
|
1.4 - 19 Nm3/h
|
VS3_15V...
|
DN15
|
20 - 200 l/h
|
1.5 - 25 Nm3/h
|
VS3_25U...
|
DN25
|
15 - 150 l/h
|
1.0 - 21 Nm3/h
|
VS3_25W...
|
DN25
|
30 - 300 l/h
|
1.5 - 35 Nm3/h
|
VS3_25A...
|
DN25
|
50 - 500 l/h
|
3 - 60 Nm3/h
|
VS3_25B...
|
DN25
|
100 - 1,000 l/h
|
6 - 100 Nm3
|
|
Warnings: Note the different ranges of precision air. The precise ranges can be found here
|
Dimensions VS3... (L = 165...200 mm) |
|
Installation and assembly indications for the PCE-VS Rotameter Rotameters should be installed vertically and free of tension. It is necessary to provide a route of entry and exit, imput of approx. 10 x DN, output of approx. 5 x DN
|
G |
DN |
L |
L1 |
L2 |
ØD |
3/4" |
10 |
165 |
171 |
199 |
35 |
1" |
15 |
185 |
191 |
223 |
43 |
1 1/2" |
25 |
200 |
206 |
250 |
60 |
|
Measuring examples with the rotameter PCE-VS |
|
Warnings: Note the different ranges of precision air. The precise ranges can be found here
|
Measuring tube material |
Nominal value |
Float |
Scale |
PA Trogamid |
DN40 / 200 ... 20,000 |
Standard |
Water l/h and % |
VS11 |
40D |
11 |
W0 |
|
Order number VS11 40D 11 W0
|
Delivery content 1 x rotameter series PCE-VS |
Additional components - Alarms contacts The alarm contacts VSK2 and VSK1 serve as external control for the flow you want to limit. They are inserted into the guide and can be adjusted to any value of the marked scale. A magnet permanently attached to the floating body opens or closes a reed switch built into the alarm contact. The switching function is stable. This means that the switching state remains active even if the magnet is removed from the contact. When installed after an alarm contact, keep in mind that you must change the standard floating body with the floating body magnet.
Voltage |
max. 230 V AC
|
Power |
max. 10 W / 12 VA
|
Current |
max. 0,5 A
|
Contact resistance |
200 mΩ
|
Isolation resistance |
1,011 mΩ
|
Ambient temperature |
0 ... 55 ºC
|
Protection |
IP65, according to DIN 40,050
|
Switchover hysteresis |
1-2 mm
|
|
|
|
|
Here you will find all the measuring instruments available at PCE Instruments.
|
Contact: PCE Instruments UK Limited
Unit 11 Southpoint Business Park
Ensign Way, Southampton
United Kingdom, SO31 4RF
Phone: +44(0) 23 809 870 30
Fax: +44(0) 23 809 870 39
|
Contact: PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8
Jupiter 33458 FL
USA
Phone: +1-410-387-7703
Fax: +1-410-387-7714
|