Air Quality Meter PCE-GA 70 Measures CO2 (carbon dioxide), temperature and humidity
|
PCE-GA 70 air quality meter is an exceptional handheld device that measures and checks air quality, and the content of carbon dioxide, humidity and temperature of interiors. This air quality meter is ideal for heating, air and climatology professionals that need this device for detecting air quality indoors, for testing air ventilation systems in schools, offices, factories and hospitals. The PCE-GA 70 air quality meter uses a sensor to define the content of carbon dioxide, air temperature and relative humidity. It can store more than 20,000 readings with a time stamp allowing to transfer data to a computer. The sensors used for testing air quality are the subject of other requirements, different to the requirements for industrial or scientific laboratory use. As such, we give priority to the properties of the sensor such as durability, low level maintenance and low energy consumption. The PCE-GA 70 air quality meter meets certain requirements with its NDIR sensor. The measurement of non-dispersed infrared (NDIR) measures accurately and has a measurement range of up to 6000 ppm. A new method of measurement (mentioned here) is developed for the analysis of air quality: the source of infrared is controlled by a brief pulse of radiation that minimises the consumption of energy. The pyro-electric detector is designed and adapted to allow the analysis of the impulse, such as a functioning pulse like that of a conventional chopper. This link shows an overview of all types of devices which will suit your needs.
|
- Large CO2 measurement range up to 6000 ppm - Triple LCD for CO2, relative humidity and air temperature - The NDIR sensor for CO2 is maintenance free - Automatic and manual data storage for up to 20,000 readings - Data Hold and memory for minimum and maximum values - Stored values can be read on the display - Alarm and time settings - RS-232 port with software included - Backlit display - Safe and easy to use
|
PCE-GA 70 air quality meter
View or print the user manual
|
Technical specifications
|
Measurement range for CO2 Accuracy Resolution
|
0 to 6000 ppm CO2 ± 3% of reading or ± 50 ppm 1 ppm
|
Temperature Accuracy Resolution
|
-20 to 60ºC ± 0.5ºC 0.1ºC
|
Air humidity Accuracy Resolution
|
10 to 95% r.h. ± 3% r.h. 0.1% r.h.
|
Manual memory
|
99 readings
|
Automatic memory
|
20,000 readings
|
Operating conditions
|
5 to 50ºC/10 to 95% r.h.
|
Power |
6 x 1.5 V batteries |
Dimensions
|
148 x 72 x 35 mm
|
Weight
|
255 g
|
Delivery content: 1 x PCE-GA 70 air quality meter, 6 x batteries, 1 x mains adaptor, 1 x software, 1 x RS-232 cable, 1 x carrying case and user manual |
|
|
Here you will find an overview of all the measuring instruments available at PCE Instruments.
|
Contact: PCE Instruments UK Limited
Unit 11 Southpoint Business Park
Ensign Way, Southampton
United Kingdom, SO31 4RF
Phone: +44(0) 23 809 870 30
Fax: +44(0) 23 809 870 39
|
Contact: PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8
Jupiter 33458 FL
USA
Phone: +1-410-387-7703
Fax: +1-410-387-7714
|