Gas Warning Station Gasmaster with Sensors of
X/TXGard-Series
Stationary gas warning station (with ATEX- permission) for adaption of gas sensors for O2, O3,
O2, O3,
CO, CO2, NH3,
H2S…
|
The gasmaster is a gas warning station/ gas warning control centre with a large display. This gas warning station is purchasable in a 1- and a 4- channel version as well as with a PC-interface. Even the gasmaster is equipped with two alarm relays, due to which e.g. signal lamps/ -horns or ventilators can be operated. The adaptable ATEX- approved gas sensors are available here as well: for flammable gasses (EX), oxygen and toxic gasses (TOX). The gas sensors are divided according to the gas, the functionality (e.g. with analogue-, relay- output…) or the display type. Thus there are sensors with an integrated display (as a local display) or without a display. There are gas sensors for nearly every gas to select: methane, propane, butane, pentane, LPG, ethanol, ethylene, hydrogen, acetylene, benzenes (unleaded), phosphine, oxygen, ozone, chlorine, carbon monoxide, carbon dioxide, hydrogen sulphide, sulphur dioxide, ammoniac, … (more gasses on request please). Systems of gas warning stations and gas warning sensors will be always delivered completely configured, calibrated and tested before delivery.
A general survey of gas meters for operator protection (mostly portable) you will find here.
If you have
further questions concerning the gas warning station, take a look at the following technical data
or call our hotline: UK customers +44(0) 23 809 870 30 / US customers +1-410-387-7703. Our
technicians and engineers will be pleased to inform you concerning the gas warning station
or any other product in terms of regulation
and control or scales
and balances of PCE
Instruments.
|
|
|
You can
see an overview of the gas
analysers for personal protection (mostly protable)
|
Technical Data about the Adaptable Sensors of Type Xgard for the Gas Station
|
Gas sensor type
|
Xgard-Typ
1
|
Xgard-Typ
5
|
TCgard
|
Images
|
|
|
|
Applicable in EX-zones
|
0,1,2
|
1,2
|
1,2
|
Detectable gasses (matching reference number for designated gas type you will find in our onlineshop) with the respective measuring range
|
0 ... 25
vol% O2
0 ... 2000 ppm H2
0 … 4 vol% H2
= (100% LFL)
0 ... 250 ppm CO
0 … 50 ppm
0 … 5 ppm CI
0 ... 1 ppm O3
0 ... 10 ppm SO2
0 …1 ppm PH3
|
0
...100 % LFL flammable gasses (methane, propane, butane, pentane, LPG, ethanol, ethylene, ...)
|
0 ... 100 vol% methane
0 ... 100 vol% CO2
|
Accuracy
|
±3 %
|
±3 %
|
±3 %
|
Display
|
- - -
|
- - -
|
- - -
|
Signal output
|
2 conductors
4 ... 20 mA
|
3 conductors
4 ... 20 mA
|
3 conductors
4 ... 20 mA
|
Alarm- /switch relays
|
- - -
|
- - -
|
- - -
|
Standardization
|
current ATEX
|
current ATEX
|
current ATEX
|
Power Supply
|
8 ... 30 V DC / 24 mA
|
8 ... 30 V DC / 24 mA
|
10 ... 30 V DC /24 mA
|
Temperature range
|
-20 ...
+50 °C
|
-40 ... +55 °C
|
+10 ... +30 °C
|
Weight
|
approx. 1 kg
|
approx. 1 kg
|
approx. 4 kg
|
Technical data about gas warning station of Xgard type (with infrared- measuring principle)
|
Gas
sensor type
|
Xgard
IR (infrared)
|
Additional equipment for Xgard IR
|
Image |
|
Applicable in EX- zone |
1.2, 21/22 |
Detectable gas (matching reference number for designated gas type you will find in our onlineshop) with according range |
0 ... 2 % CO2
0 ... 5 % CO2
(depending on sensor)
0 ...100 % LFL flammable gasses (methane, propane, butane, pentane, LPG, ethanol, ethylene, ...) |
Accuracy |
±2 % (whole range) |
Display |
---- |
Signal output |
4 ... 20 mA (3-wire) |
Alarm- / Switch- Relays |
--- |
Standardisation |
current ATEX/ UL |
Protection |
IP65/IP66 with waterproof cap |
Power Supply |
10 ... 30 V DC / 50 mA |
Temperature
range |
-20 ... +50 °C |
Measures |
156 x 166 x 111 mm |
Weight |
1 kg (aluminium- enclosure) |
Further gas warning stations/ gas sensors for CO2 you will find
here. |
|
|
|
Technical
Data about Adaptable Sensors Type TXgard for the Gas Warning Station
|
Gas sensor type
|
TXgard-IS+
|
TXgard-Plus
|
Flamgard-Plus
|
Images
|
|
|
|
Applicable in EX-zone
|
0, 1, 2
|
1, 2
|
1, 2
|
Detectable gasses (matching reference number for designated gas type you will find in our online shop) with according range.
|
0 …25 vol% O2
0 … 500 ppm CO
0 ... 50 ppm H2S
0 … 1 ppm O3
0 … 2 ppm PH3
|
0 ... 25 vol% O2
0 ... 50 ppm H2S
0 … 500 ppm CO
0 … 50 ppm NH3
|
0 ...100 % LFL flammable gasses (hydrogen, methane, propane, butane, pentane, ethanol, ethy-
lene, ...)
|
Accuracy
|
±3 %
|
±3 %
|
±3 %
|
Display
|
yes
|
yes
|
yes
|
Signal output
|
2 conductors
4….20 mA
|
3 conductors
4….20 mA
|
3 conductors
4…. 20 mA
|
Alarm/ Switch relays
|
---
|
yes
|
yes
|
Standardisation
|
current ATEX
|
current ATEX
|
current ATEX
|
Power Supply
|
11 ... 30 V DC /
20 mA
|
10 ... 30 V DC /
210 mA
|
10 ... 30 V DC /
210 mA
|
Temperature range
|
-20 ... +50 °C
|
-10 ... +55 °C
|
-10 ... +55 °C
|
Weight
|
0.7 kg
|
2.2 kg
|
2.2 kg
|
Technical
Data for Gas Warning Station Gasmaster (for 1-4 Gas Sensors)
|
Digital display and control unit for gas sensors shown above. The gas warning station gasmaster is deliverable in two models. GM-1 for an Input sensor and GM-4 for up to 4 gas sensors (simultaneous display of all 4 measuring datas). Those two types are besides the standard analogue output even purchasable with an additional RS-232-data interface. Furthermore the device is equipped with relay outputs e.g. for activation of an all-round warning light. The enclosure is dust- and waterjet- proof. The gas
warning station is further equipped with a standard backup-battery in order to operate even during power outage.
|
|
Model
|
GM-1
|
GM-4
|
Sensor input
|
1 input for one gas sensor
|
4 inputs for 4 gas sensors
|
Output
|
- relay output (250 VAC / 8 A),
- 1 analogue output (4 ... 20 mA),
- 1 digital RS-485,
-1 optional digital RS-232
|
- 4 relay outputs (250 VAC / 8 A),
- 4 analogue outputs (4…20 mA),
-1 digital RS-485,
-1 optional digital RS-232
|
Data storage
|
For 300 processes (alarms, system errors, settings alternation)
|
Alarm indicators
|
- Digital display for a gas in ppb, ppm or %,
- background lightening, - LEDs for alarm, ready-to-operate and error
|
- Digital display for
4 gasses in ppb, ppm or %,
- background lightening, - LEDs for alarm, ready-to-operate and error
|
Alarm indicators
|
acoustic: 85 dB
beep
visual: Alarm-LEDs
|
Operating temperature range
|
-10 ... +50 °C
|
Operating humidity range
|
0 ... 95 % RH (non condensing)
|
Power supply
|
90 ... 264 VAC / 50 ... 60 Hz
|
Backup battery
|
1.2 Ah
|
Enclosure material
|
frame with stainless steel/ front with ABS
|
Measures
|
288 x 278 x 10 mm
|
Weight
|
4.5 kg
|
Standardisation
|
EN 50270 (EMC), EN 61010-1 low voltage
|
Protection
|
IP 54
|
Delivery
Content
|
Gas warning station gasmaster (GM-1 or GM-4): device with instruction manual
Sensors: all gas sensors are delivered calibrated, including calibration paper and operating instruction
|
Optional Equipment
|
-
Re-calibration of gas sensors according to DIN EN ISO including test certificate
(either in case of
transmittal
of sensors to us or directly on-site through one of our service-partners)
- RS-232- interface for gasmaster (GM-1 or GM-4)
- PC-software including RS-232-USB-adapter (in case of using the optional
RS-232-interface)
- RS-232-USB adapter for optional RS-232-interface for gasmaster (GM-1 and GM-4)
- Zener barrier for sensor types Xgard type 1 and TXgard-IS+ (for Ex-zones 0
mandatory (once
per detector))
- Signal-horn 230 VAC/ 90 dB
|
|
|
Here you will find an overview of all the measuring instruments available at PCE Instruments. |
Contact: PCE Instruments UK Limited
Unit 11 Southpoint Business Park
Ensign Way, Southampton
United Kingdom, SO31 4RF
Phone: +44(0) 23 809 870 30
Fax: +44(0) 23 809 870 39
|
Contact: PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8
Jupiter 33458 FL
USA
Phone: +1-410-387-7703
Fax: +1-410-387-7714
|
Below you will find an overview of the different product groups of measuring instruments
|
|
|
|
|
Below you will find an overview of the different
product groups of scales and scales |
|
|
|
|
|