PAT Tester PKT-2765
Pat tester according to Pat-Norm 0701/702 and Pat-Norm 0751 for electromedical meters
|
Pat
testers for patients-discharge-current according to PAT 0751. You have to do different measurements on
electric meters of the protection classes I, II and III for the protection of users if you do changes or new installations. With this
Pat tester the measurements according to DIN
PAT 0701/ 0702/ 0751 and VBG4 can be carried and the control of protective conductor resistance, surrogate-arrester-, current-current, insulation resistance, voltage-freedom and earth leakage current is given. Gauges- and patient-discharge-currents can be measured in residual-current-procedure or as surrogate-arrester-current. Take care to the implementation: In the
surrogate-patients-discharge-current every application partial has to be measured particularly. The RS-232 interface and the software enable the documentation of the measurements and there evaluation.
If you have
further questions concerning the Pat
tester, take a look at the following technical data
or call our hotline: UK customers +44(0) 23 809 870 30 / US customers +1-410-387-7703. Our
technicians and engineers will be pleased to inform you concerning the Pat tester
or any other product in terms of regulation
and control or scales
and balances of PCE
Instruments.
|
|
|
- limit display (3 red LED’s)
- fast PLL-controlled LCD-display
- calibrated measurement line in the region R-PE
- safety: TÜV/GS; IEC-1010-1
- complies the PET norm 0701/ 0702 /0751
|
- provided supplies: testing cable set with test
probe, cable with alligator clamp for
measurements of the protective conductor
resistance and software, interface cable and
manual
|
Technical
data
of the Pat Tester PKT-2765
|
Protective conductor resistance
|
Range
|
R-PE 20 ... 1999 mΩ
|
Method
|
UO ~ 20 VAC, IK > 200 mA
|
Limit display
|
100/ 300/ 1000 mΩ
|
Isolation resistance
|
Range
|
R-ISO 0 ... 19,99 MΩ
|
Method
|
UN > 600 V DC, IK < 5 mA
|
UN > 600 V DC, IK < 5 mA
|
2,00 / 1,00 / 0,30 MΩ
|
Surrogate-arrester-current
|
Range
|
I-EA 0 ... 19,99 mA
|
Method
|
UO ~ 230 VAC, IK < 3,5 mA
|
Limit display
|
0,50/ 1,00/ 3,50 mA
|
Arrester-current 1
|
Range
|
I-ABL 10 ... 1999 μA
|
Method direkt
|
direkt
|
Limit display
|
0,50/ 100/ 500 μA
|
Arrester-current 2
|
Range
|
I-ABL 2 ... 19,99 mA
|
Method direkt
|
direkt
|
Limit display
|
2,50/ 3,50/ 5,00
mA
|
Operating voltage
|
230 V AC/50 Hz
|
Dimensions
|
230 V AC/50 Hz
|
Weight
|
500 g
|
Norm
|
TÜV/GS; IEC-1010-1; VDE 0701/ 0702/ 0751
|
Delivery
Content
|
1 x portable
Pat tester PKT-2765, 1 x testing-cable-set, 1x cable with alligator clamp for measurements of the protection conductor resistance, software, interface cable and manual
|
Additional
|
- yearly / rotational
Pat Tester check
- ISO-calibration-certificate
|
|
|
Here you will find an overview of all the measuring instruments available at PCE Instruments. |
Contact: PCE Instruments UK Limited
Unit 11 Southpoint Business Park
Ensign Way, Southampton
United Kingdom, SO31 4RF
Phone: +44(0) 23 809 870 30
Fax: +44(0) 23 809 870 39
|
Contact: PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8
Jupiter 33458 FL
USA
Phone: +1-410-387-7703
Fax: +1-410-387-7714
|